Search
EN

Horizontal and vertical laminar flow chambers

Laminar flow chambers are designed to restrict an enclosed workspace in the laboratory with forced air circulation inside. The airflow passing through a HEPA filter protects the operator and the working sample from contamination. There are horizontal laminar flow chambers (direction of filtered air horizontally, ensuring protection of the biological materials handled) and vertical laminar flow (direction of filtered air downwards vertically, guaranteeing operator protection against possible risk contaminants). Equipped with long-lasting white light lamps and UV lamps for sterilization.
SKU: BIOAL-H180

Register or Log In in order to request a quote

Register Login

Products specifications
Sizes
Laminar flow chambers are designed to restrict an enclosed workspace in the laboratory with forced air circulation inside. The airflow passing through a HEPA filter protects the operator and the working sample from contamination. There are horizontal laminar flow chambers (direction of filtered air horizontally, ensuring protection of the biological materials handled) and vertical laminar flow (direction of filtered air downwards vertically, guaranteeing operator protection against possible risk contaminants). Equipped with long-lasting white light lamps and UV lamps for sterilization.


HORIZONTAL LAMINAR FLOW CHAMBERS
Horizontal laminar flow chambers with manual or automatic operation. Polycarbonate front panel and stainless steel work surface. They have an LCD control screen and include white light lamp, UV sterilization lamp (30W) and power socket inside. Optical and audible filter clogging alarm. "Autocheck" and error detection system. Microprocessor control of air flow, lighting and fan operating hours counter.

Specifications:
- Sterilization with UV lamp (30W) and white light for lighting;
- Horizontal laminar flow for sample protection.
- Airflow: 100% extraction
- Horizontal laminar air flow velocity from 0.25 to 0.50 m/s;
-Main filter HEPA Absolute EN 1800, with efficiency of 99.997% in particles of 0.3 µm;
- Pre-filter Class EU3 (G3) according to PN-EN779;
- Noise level lower than 57 dB;

Accessories available (not included) such as: table to support the camera, power socket, vacuum valve.
- Dimensions of the work surface: 179 x 63 x 60 cm (there are models with different dimensions);
- Exterior dimensions (WxDxH): 185 x 79 x 95 cm.


VERTICAL LAMINAR FLOW CHAMBERS
Vertical laminar flow chambers with manual or automatic operation. Full front panel with half-height opening via hinges, stainless steel work surface, side walls and rear safety glass. They have an LCD control screen and include white light bulb, and power socket inside on some models. Optical and audible filter clogging alarm. "Autocheck" and error detection system. Microprocessor control of air flow, lighting and fan operating hours counter.

Specifications:
- Sterilization with UV lamp (30W) and white light for lighting;
- Vertical laminar flow.
- Airflow: 100% extraction
- Vertical laminar air flow velocity of 0.35 m/s;
-Main filter HEPA Absolute EN 1822, with efficiency of 99.995% in particles of 0.3 µm;
- Pre-filter Class EU3 (G3) according to PN-EN779;
- Noise level lower than 55 dB;
- Current socket (available on some models);
- Available accessories (not included) such as: table to support the camera, vacuum valve.
- Dimensions of the work surface: 159 x 62 x 50 cm (there are models with different dimensions);
- Exterior dimensions (WxDxH): 161 x 64 x 78 cm.


VERTICAL LAMINAR FLOW CHAMBERS
Vertical laminar flow cabinets ideal for handling non-pathogenic samples, cell and tissue cultures. High efficiency fan with operating hours counter and partial recirculation of filtered air, for longer life of HEPA filters and pre-filters. They have digital control by microprocessor and two power sockets.

Specifications:
- Sterilization with UV lamp (30W) and white light for lighting;
- 15 min UV light timer;
- Vertical laminar air flow velocity of 0.40 ±20 % m/s;
- HEPA main filter with 99.999% efficiency on particles larger or equal to 0.3 µm (class H14 according to EN1822);
- Pre-filter with an efficiency of 85% for particles larger or equal to 0.5 µm (class G14 according to EN779);
- Noise level lower than 60 dB;
- Available accessories (not included) such as: table to support the camera, vacuum valve and gas valve, front closure with UV lamp, HEPA filter and pre-filters;
- Dimensions of the work surface: 119.5 x 63 x 65 cm (there are models with different dimensions);
- Exterior dimensions (WxDxH): 136 × 79.5 × 140 cm.
Similar Products

Class II biological safety chambers

Class II biological safety chambers are ideal laminar flow chambers to protect the operator, the product and the environment. The cabin operates with negative pressure, preventing the release of contaminated air into the environment. Designed to offer total protection, the air circulating inside the chamber is filtered by a HEPA filter (99.999% efficiency), and the fresh air is also filtered through another HEPA filter. They also have an air speed compensation system before the resistance increases by the filter and alarms in case of electrical supply failures. These cameras are used in different areas of research, such as microbiology, virology and genetics and in the various service laboratories, where moderate biological risk samples are handled.

PCR UV chambers

Ozone-free, high-density UV decontamination PCR chambers with long-life lamps and UV lamps that automatically turn off when the protection panel is opened. UV irradiation disinfects the work area by inactivating DNA/RNA fragments during a 15 to 30 minute exposure. The cameras have low noise and low power consumption.